Statler Configuration

Suggested voltages current as of 11 March 2013.

Reservations

Reservations can be made online by following this link.

You will need an approved USER account. Apply for one here.

Statler Configuration: 

488 Laser  || 633 Laser  || 407 Laser

488 LASER:

Detector

LP

Band Pass

Fluor's

Suggested Voltage

A

735

780/60

PE-Cy7

700

B

685

695/40

PerCp-Cy5.5, PE-Cy5.5

650

C

655

670/14

7AAD, PE-CY5, Cychrome, PerCP

600

D

595

610/20

PI, PE-Texas Red, PE-Alexa610

600

E

556

576/30

PE

650

F

525

530/30

YFP, FITC, Alexa 430, Alexa 488, DCF, TO

600

G

502

510/20

GFP

600

H

-

488/10

SSC

 

633 LASER

Detector

LP

Band Pass

Fluor's

Suggested Voltage

A

735

780/60

APC-Cy7, APC-Alexa750

600

B

685

720/40

APC-Cy5.5, APC-Alexa700

500

C

--

660/20

APC, Alexa 660, To-Pro3

500

407 LASER:

Detector

LP

Band Pass

Fluor's

Suggested Voltage

A

502

530/30

Alexa 430, Qdot520, Pacific Orange

500

B

470

475/22

CFP, Hoescht

400

C

---

450/40

Alexa 405, DAPI, Cascade Blue, Pacific Blue

350

 

Instrument Locations

  • Statler - 3-4128
  • Waldorf - 3-4138