Field Test of MALDI

MALDI Field Test Results

  • A dilution series of seven proteins were analyzed
  • 1000 to 1 ng of protein were loaded in half-log dilution
Protein Mass (kDa) pmoles of 1ug band pmoles of 0.3 ug bands Field Test gel
ß-gal 116 9 3
Albumin, serum 66 15 5
Glut.Dehydrogenase 61 16 5
peroxidase 44 23 7
GPDH 36 26 9
superoxide dismutase 16 62 20
α-Lactalbumin 14 71 22
MALDI-Performance:
  • MALDI-TOF analysis of all protein bands ( > 312 ng ) produced excellent quality and publishable protein identifications.
  • 312 ng is not our limit of detection; however, this is what we currently guarantee by MALDI.

Alternative Mass Spectrometry Methods:

For increased coverage and sensitivity, samples can be simultaneously applied to our:

  • Nanospray LC-MS/MS instruments. (Method # 40, 50)
  • Integrated HPLC, robot-spotter, high-throughput MALDI-TOF/TOF service (LC-MALDI, Method # 30)