MRI contrast and imaging parameters


  Contrast: Difference between different regions

  

  TR: repetition time, affect longitudinal magnetization.
  TE: echo time, affect transverse magnetization.
  In general,

  ==> Long TR, short TE --> density-weighted
  Short TR, short TE --> T1-weighted
  Long TR, long TE --> T2-weighted
  Different tissues ~ different environment for protons detected.
  T1, T2 are field dependent. At B0 = 1.5-2 T,
  T1: 100ms (fat), 500ms (liver), 1000ms (brain), 2000ms (CSF)
  T2: 20ms (muscle), 50ms (liver), 100ms (brain), 500ms (CSF)
  FOV (field-of-view): spatial coverage of image in each direction
  resolution: number of points in image (typically 256x256)

[chapter top] [next chapter]