Anesthesia Machine/Vaporizer Preventive Maintenance PolicyPurpose: Electrical Safety inspection and battery replacement for Desflurane Vaporizers. Physical inspection for loose or missing screws, sharp edges, or loose connectors. Visual inspection for mechanical or electrical damage. Verify operation of safety interlock if applicable. Verify operation of handwheel lock and release button. Verify handwheel turns smoothly through entire range. Check Fill / Drain system operates properly and that there are no leaks. Verify concentration delivery accuracy at 0, 1, 3, & 5% settings using room temperature 100% Oxygen at 5 LPM flow is within manufacturer’s specifications. References: Updated 5/2/06 |
|||