Dr. Teeth Configuration

Suggested voltages current as of Oct 2, 2017.

Detailed information about this instrument can be found on the baseline report (available upon request).

Reservations

Reservations can be made online by following this link.

You will need an approved USER account. Apply for one here.

To design a panel using Fluorofinder please click here.

LSR-II Configuration:

BLUE 488 LASER (50mW):

Detector

LP

Band Pass

Effective Ranges (nm)

Fluor's

Low Starting Voltage

BLUE A

685

710/50

685-735

PerCP-Cy5.5

550

BLUE B

505

530/30

515-545

BB 515

450

BLUE C

  488/10   SSC  

FSC

 

 

 

 

 

RED 640 LASER (40mW):

Detector

LP

Band Pass

Effective Range (nm)

Fluor's

Low Starting Voltage

RED A

750

780/60

750-810

APC-Cy7

500

RED B

690 730/45 707.5-750 APC-R700 450

RED C

 

670/14

663-677

APC

450

VIOLET 405 LASER (50mW):

Detector

LP

Band Pass

Effective Range (nm)

Fluor's

Low Starting Voltage

VIOLET A

750

780/60

750-810

BV 786

600

VIOLET B

690

710/50

690-735

BV 711

550

VIOLET C

630

660/20

650-670

BV 650

500
VIOLET D 600 610/20 600-620 BV 605 500
VIOLET E 505 525/50 505-550 BV 480 400
VIOLET F   431/28 417-445 BV 421 350

YELLOW-GREEN 561 LASER (50mW):

Detector

LP

Band Pass

Effective Range (nm)

Fluor's

Low Starting Voltage

YG A

750

780/60

750-810

PE-Cy7

550

YG B

685

710/50

685-735

PE-Cy5.5

450

YG C

635 670/30 655-685 PE-Cy5 450

YG D

600

610/20

600-620

PE-Texas Red

400

YG E

 

586/15

578.5-593.5

PE, PI, Alexa 532

400

Ultra Violet 355 LASER (20mW):

Detector

LP

Band Pass

Effective Range (nm)

Fluor's

Low Starting Voltage

UV A

690

740/35

722.5-757.5

BUV 737

600

UV B

 

379/28

365-393

BUV 395

450