LSR Fortessa Configuration: Low starting voltages posted at the machine

BLUE 488 LASER (50mW):

Detector

LP

Band Pass

Effective Ranges (nm)

Fluors 

(check Fluorofinder 
for more info)

BLUE A

685

710/50

685-735

PerCP-Cy5.5

BLUE B

505

530/30

515-545

BB 515

BLUE C

 488/10 SSC

FSC

 

 

 

 

RED 640 LASER (40mW): 

Detector

LP

Band Pass

Effective Range (nm)

Fluors 

(check Fluorofinder 
for more info)


RED A

750

780/60

750-810

APC-Cy7

RED B

690730/45707.5-750APC-R700

RED C

 

670/14

663-677

APC

VIOLET 405 LASER (50mW): 

Detector

LP

Band Pass

Effective Range (nm)

Fluors 

(check Fluorofinder 
for more info)


VIOLET A

750

780/60

750-810

BV 786

VIOLET B

690

710/50

690-735

BV 711

VIOLET C

630

660/20

650-670

BV 650

VIOLET D600610/20600-620BV 605
VIOLET E505525/50505-550BV 480
VIOLET F 431/28417-445BV 421

YELLOW-GREEN 561 LASER (50mW): 

Detector

LP

Band Pass

Effective Range (nm)

Fluors 

(check Fluorofinder 
for more info)


YG A

750

780/60

750-810

PE-Cy7

YG B

685

710/50

685-735

PE-Cy5.5

YG C

635670/30655-685PE-Cy5

YG D

600

610/20

600-620

PE-Texas Red

YG E

 

586/15

578.5-593.5

PE, PI, Alexa 532

Ultra Violet 355 LASER (20mW):

Detector

LP

Band Pass

Effective Range (nm)

Fluors 

(check Fluorofinder 
for more info)

UV A

690

740/35

722.5-757.5

BUV 737

UV B

 

379/28

365-393

BUV 395